Command Reference
R&S
®
ZNB/ZNBT
772
User Manual 1173.9163.02 ─ 38
CALCulate<Ch>:PARameter:SELect
<TraceName>
Selects an existing trace as the active trace of the channel. All trace commands with-
out explicit reference to the trace name act on the active trace (e.g.
).
CALCulate<Ch>:PARameter:SELect
is also neces-
sary if the active trace of a channel has been deleted.
Suffix:
<Ch>
.
Channel number.
Parameters:
<TraceName>
Trace name, e.g. 'Trc4'. See "Rules for trace names" in
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named
Ch4Tr1
to measure the
input reflection coefficient S
11
. The trace is the active trace in
channel 4.
CALC4:PAR:SDEF 'Ch4Tr2', 'S22'
Create another trace named
Ch4Tr2
to measure the output
reflection coefficient S
22
. Again this new trace becomes the
active trace in channel 4.
CALC4:PAR:SEL 'Ch4Tr1'
Select the first trace
Ch4Tr1
as the active trace.
CALC4:FORM MLIN
Calculate the magnitude of S
11
and display it in a linearly scaled
Cartesian diagram.
7.3.2.13
CALCulate:PHOLd...
The
CALCulate:PHOLd...
commands control the max hold (peak hold) function.
CALCulate<Chn>:PHOLd
<HoldFunc>
Enables, disables, or restarts the max hold and the min hold functions.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HoldFunc>
MIN | MAX | OFF
MIN - Enable the min hold function.
MAX - enable the max hold function.
OFF - disable the max hold or min hold function.
*RST:
OFF
Example:
*RST; :CALC:PHOL MAX
Reset the instrument and enable the max hold function.
CALC:PHOL OFF; PHOL MAX
Restart max hold.
Manual operation:
See
SCPI Command Reference