GUI Reference
R&S
®
ZNB/ZNBT
612
User Manual 1173.9163.02 ─ 38
Remote command:
MMEMory:STORe:TRACe:OPTion:TABS
Trim Leading Whitespace ← Touchstone Export Options
If checked, whitespace at the beginning of each line is removed
Remote command:
MMEMory:STORe:TRACe:OPTion:TRIM
Positive Number Prefix ← Touchstone Export Options
Positive numbers can either be prefixed by blanks, by plus signs or not at all
Remote command:
MMEMory:STORe:TRACe:OPTion:PLUS
Skip Separator Lines ← Touchstone Export Options
If checked, the content parts are no longer separated by blank lines
Remote command:
MMEMory:STORe:TRACe:OPTion:SSEParator
Power Tab
The settings in this tab define how the VNA sets the output power between sweeps.
This applies to all sweep modes but is particularly useful in single sweep mode.
Keep
Auto
Reduce
Power
Time
Settling Delay
End Measurement
Restart Sweep
Figure 5-56: Power control at sweep end in Single Sweep mode (power sweep)
By default, at sweep end the output power of the first measurement point is restored
("Auto" mode). After a small settling time (~10 ms) the sweep can be restarted.
Selecting a different behavior may result in a significantly longer total measurement
time, in particular if a long "Settling Delay" is used. Don't forget to revert the
to "Auto" mode if is not necessary to "Reduce" or "Keep" the
power at the sweep end. A preset is not sufficient for this purpose!
Setup Softtool