GUI Reference
R&S
®
ZNB/ZNBT
266
User Manual 1173.9163.02 ─ 38
If auto-detection fails because of a high attenuation in the signal path, you can either
enter the port assignment manually or connect matching port numbers and select
"Default Port Assignment".
Remote command:
[SENSe:]CORRection:COLLect:AUTO:PORTs:CONNection?
Default Port Assignment
Restores the default port assignment(s).
Remote command:
Start
.
If the configured port assignments are invalid, this action is disabled.
Step 3: Cal Unit
During the calibration phase the R&S
ZNB/ZNBT displays a "Cal Unit" screen that
guides the user through the actual correction data acquisition.
In each calibration step
●
the calibration unit has to be (re-)connected according to the depicted port assign-
ment; auto-detection is possible
●
the related test ports are calibrated with the "Calibration Type" selected in
●
a subsweep is performed for every required test port (pair), for every possible path
(if external switch matrices are involved) and for every required standard
After all these calibration steps have been completed, the resulting system error cor-
rection can be calculated and applied.
Cal Softtool