Operating Manual
Page 52 / 116
Copyright © REA Elektronik
REA ScanCheck 3n
V 1.2.0 24/02/2020
REA
VERIFIER
Other measured data and threshold values always shown in the reference test report in the
context must be ignored for this test.
Measuring tolerances for apertures from 6 to 10mil
Aperture
1
6 mil, 0.15 mm
Light source
670 nm at 45°
Symbol
Passed 4.0
Code type
REA test code
Code content
4008654123454
The following table lists the criteria to be verified in accordance with ISO/IEC 15426-1.
Measuring parameters
Unit
Tolerance
Reference
Rmax (see histogram)
2
%
+/- 5
Measurement
values on the
stamped
reference test
report
Rmin (see histogram)
3
%
+/- 3
Decodability
%*
+/- 8
Defects
%*
+/- 8
*In standard ISO/IEC 15426-1 the values for defects and decodability are cited as absolute values
with reliable tolerances of +/-0.08. For improved comprehensibility, we refer to these two values on
the test reports as percentages. The corresponding tolerance values are the/- 8. In the
event that the tolerances for decodability and/or defects are exceeded, the device must be
inspected and serviced at the manufacturer.
Other test criteria
Measuring
parameters
Unit
Tolerance
Reference
Decode
-
No tolerance
Measurement
values on the
stamped
reference test
report
Symbol contrast
%
+/- 8
Modulation
%
+/- 11
Module size (Z module)
µm
+/- 5
Average bar variance
(shown as mean)
µm
+/- 5
Maximum bar variance
(shown as bar, two values
must be checked)
µm
+/- 15
Tolerance refers
to the mean
value on the
reference test
report
E-Value variance
µm
+/- 20
Tolerance refers
to 0
The terms “calibration” and “adjustment” are used in accordance with the definition of DIN 1319-1
and JCGM 200.
1
Aperture – the light point size in accordance with the definition of ISO/IEC 15416
2
Rmax – the maximum reflection in the symbol determined in accordance with ISO/IEC 15416
3
Rmin – the minimum reflection determined in accordance with ISO/IEC 15416