APPLIED OPERATIONS
/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2 - 1 4
Page
2-2-1-3 Setting the Observation Condition
1 Setting the Objective Ma gnification
1. From the drop-down list on the center of the [Acquire] panel, select the objective
being used with the microscope.
If the magnification of the objective in use and that set here do not
match, the measurement results will be inaccurate.
2 Setting the Zoom Ratio to 1X
1.
Use the [Zoom] scale to set the zoom ratio to “X1”.
Fig. 2-6 [Pan]/[Zoom] Group Box
3 Setting the Channels
1.
In the Channel 1 group box, check the check box showing the applicable dyeing
method to make the image acquisition ready.
2.
In the Channel 2 group box, check the check box showing the applicable dyeing
method to ready the image acquisition.
3. In case of transmitted light observation, also make sure that the check box
showing the Ch3 dying method is check-marked to indicate that Ch3 is ready for
image acquisition.
NOTE
CH2 check box
CH1 check box
CH3 check box
Содержание FLUOVIEW FV300
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 26: ......
Страница 28: ......
Страница 38: ......
Страница 114: ......
Страница 291: ...APPLIED OPERATIONS Protocol processor IV OPERATION INSTRUCTIONS IV 2 1 7 7 Page Fig 2 54 Panel After a Repeated Scan...
Страница 422: ...APPLIED OPERATIONS Viewing 3D Image IV OPERATION INSTRUCTIONS IV 2 3 0 8 Page Fig 2 133 Panel Showing Stereo 3D Images...
Страница 486: ......
Страница 506: ......
Страница 512: ......
Страница 518: ......
Страница 520: ......
Страница 548: ......
Страница 552: ......
Страница 554: ......
Страница 556: ......
Страница 562: ......
Страница 563: ......