Getting Started FLUOVIEW/
Online Help
IV.
OPERATION INSTRUCTIONS
IV.
1 - 6 7
Page
z
Clicking the <Link Setting> button displays the [Link Setting] dialog box as shown
below.
Click the desired check box to be checked.
Select the <Close> button to close the dialog box.
The microscope settings are automatically saved and read out when the software is
started up next time.
Checking here links the objective in the [Nosepiece] group box
with the condenser turret.
Checking here escapes the stage or revolving nosepiece when
the objective is selected and changed in the [Nosepiece] group
box
Checking here disengages the top lens from the light path
when the objective of
u
4 or lower magnification is selected in
the [Nosepiece] group box. (BX only)
Checking here closes the FL shutter and engages Analyzer
and Polarizer into the light path when the TD lamp is set to ON
in the [Lamp] group box.
Checking here sets the TD lamp to OFF when the fluorescent
cube is selected in the [Mirror Unit].
Checking here disengages the top lens from the light path
when the fluorescent cube is selected in the [Mirror Unit].
Checking here enables parfocality correction when the
objective is selected and changed in the [NosePiece] group
box.
<Import Setting> button
Loads the microscope
settings already
registered to reflect it.
<Export
Setting>button
Saves the current
Содержание FLUOVIEW FV300
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 26: ......
Страница 28: ......
Страница 38: ......
Страница 114: ......
Страница 291: ...APPLIED OPERATIONS Protocol processor IV OPERATION INSTRUCTIONS IV 2 1 7 7 Page Fig 2 54 Panel After a Repeated Scan...
Страница 422: ...APPLIED OPERATIONS Viewing 3D Image IV OPERATION INSTRUCTIONS IV 2 3 0 8 Page Fig 2 133 Panel Showing Stereo 3D Images...
Страница 486: ......
Страница 506: ......
Страница 512: ......
Страница 518: ......
Страница 520: ......
Страница 548: ......
Страница 552: ......
Страница 554: ......
Страница 556: ......
Страница 562: ......
Страница 563: ......