Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
IV.
OPERATION INSTRUCTIONS
IV.
1 - 4 7
Page
1-2-12-5 Setting the XY Observation Mode
1. In the [Mode] group box in the [Settings] sub-panel, select the [Surface] option
button.
2. In the [Mode] group box in the [Acquire] panel, select [800 by 600] from the [Size]
drop-down list.
3.
On the center of the [Acquire] panel, select the XY observation mode option button.
1-2-12-6 Repeated Scanning Operation
1. Select the <XY Repeat> button. The acquired image will be displayed in the [Live]
panel.
Use the <FOCUS> button to acquire image at an even higher speed. If
the specimen is already being scanned, stop scanning with the <STOP
SCAN> button before selecting the <XY Repeat> button.
The Focus function reduces the scanning time by line skipped scan.
As a result, the acquired images become coarse.
Do not move FLOUVIEW FV300 Menu while acquiring an image.
<XY Repeat> button
<Focus> button
NOTE
TIP
NOTE
Содержание FLUOVIEW FV300
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 26: ......
Страница 28: ......
Страница 38: ......
Страница 114: ......
Страница 291: ...APPLIED OPERATIONS Protocol processor IV OPERATION INSTRUCTIONS IV 2 1 7 7 Page Fig 2 54 Panel After a Repeated Scan...
Страница 422: ...APPLIED OPERATIONS Viewing 3D Image IV OPERATION INSTRUCTIONS IV 2 3 0 8 Page Fig 2 133 Panel Showing Stereo 3D Images...
Страница 486: ......
Страница 506: ......
Страница 512: ......
Страница 518: ......
Страница 520: ......
Страница 548: ......
Страница 552: ......
Страница 554: ......
Страница 556: ......
Страница 562: ......
Страница 563: ......