APPLIED OPERATIONS
/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2 - 1 9
Page
9 Adjusting the image brig htness
1. While observing the image in the [Live] panel, change the setting values of the
PMT voltage
[PMT],
Offset
[Offset] and
Gain
[Gain] in the [Acquire] panel. The
functions of these parameters are as described below.
PMT Voltage
Increasing this value improves the sensitivity.
However, too high a sensitivity makes image noise noticeable.
If sufficient brightness cannot be obtained by setting the PMT voltage to
800 V, leave it as it is and increase Gain. This will usually provide a better
result than using a PMT voltage over 800 V.
Offset
Darkens the image at the ratio set during image acquisition.
(This value can be set independently from the Gain.)
Gain
Brightens the image at the ratio set during image acquisition.
(This value can be set independently from the Offset.)
For example...
When the acquired image is dark or the observation targets are hardly visible, increase
PMT voltage
to improve the sensitivity. If the image resulting from this becomes too
bright, first try decreasing
PMT voltage
slightly. If the image background is still too
bright, increase
Offset
to darken the background. As this also reduces the brightness of
the observation target, increase
Gain
as required so that the observation target is
clearly visible.
Gamma Correction provides you more brightened image even if it was acquired with
darkness.
Refer to “2-5-2-2 LUT Graph Editing by Gamma Correction “ for details in this operation.
Clicking this field allows the value to
be changed on a large scale.
The ND value which are usually used
are displayed in green.
Clicking this button allows fine
adjustment of the value.
Clicking the <+> or <-> button or the
field displays the set value. The set
value can be varied by direct input
from the keyboard.
Содержание FLUOVIEW FV300
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 26: ......
Страница 28: ......
Страница 38: ......
Страница 114: ......
Страница 291: ...APPLIED OPERATIONS Protocol processor IV OPERATION INSTRUCTIONS IV 2 1 7 7 Page Fig 2 54 Panel After a Repeated Scan...
Страница 422: ...APPLIED OPERATIONS Viewing 3D Image IV OPERATION INSTRUCTIONS IV 2 3 0 8 Page Fig 2 133 Panel Showing Stereo 3D Images...
Страница 486: ......
Страница 506: ......
Страница 512: ......
Страница 518: ......
Страница 520: ......
Страница 548: ......
Страница 552: ......
Страница 554: ......
Страница 556: ......
Страница 562: ......
Страница 563: ......