APPLIED OPERATIONS
/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2 - 8 7
Page
The speed recommended for point scan are as follows.
2
P
s(The scroll bar indicates “Fast” position. It is useful for the
specimen changes rapidly.), 10
P
s, 100
P
s(The scroll bar indicates
“Slow” position. It is useful for the specimen changes slowly.)
5.
Set the time for measurement.
From the page tabs on the bottom right of the [Acuire] panel, select the [Time
Series] sub-panel.
Set the time for measurement with the <
> or <
> button in the [Scanning] text
box.
6.
Set the starting method for image acquisition.
(If you have FLUOVIEW TIME COURSE software)
For setting procedure, see “FLUOVIEW TIME COURSE software user’s manual”.
7.
Click the <Pt
x
T> button to acquire the image.
The image at the coordinates set in the [Live] panel appears in the image display
area and the graph showing the fluorescence intensity also appears in the graph
window.
TIP
[Scanning] text box
Содержание FLUOVIEW FV300
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 26: ......
Страница 28: ......
Страница 38: ......
Страница 114: ......
Страница 291: ...APPLIED OPERATIONS Protocol processor IV OPERATION INSTRUCTIONS IV 2 1 7 7 Page Fig 2 54 Panel After a Repeated Scan...
Страница 422: ...APPLIED OPERATIONS Viewing 3D Image IV OPERATION INSTRUCTIONS IV 2 3 0 8 Page Fig 2 133 Panel Showing Stereo 3D Images...
Страница 486: ......
Страница 506: ......
Страница 512: ......
Страница 518: ......
Страница 520: ......
Страница 548: ......
Страница 552: ......
Страница 554: ......
Страница 556: ......
Страница 562: ......
Страница 563: ......