Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
IV.
OPERATION INSTRUCTIONS
IV
. 1 - 2 2
Page
1-2-1 Turning Power On
Set the power switch of each unit to ON.
When it is necessary only to read data, it is not necessary to turn the laser power supply
and reflected light power supply units ON.
To stabilize the laser beam output, we recommend leaving the system for warm-up after
turning the laser power ON. The warm-up period should be more than 10 minutes when
the laser system in use uses an Ar laser or more than 30 minutes when it uses a Kr laser
or Green HeNe laser.
x
Turning the computer (PC) ON
When the power outlet unit is used, turning the computer power ON
also turns ON the power supplies to the microscope, monitor, power
supply unit, hard-copy unit and transmitted light unit (when the
transmitted light detector is used).
x
Turning the power unit ON
(1) Turn the key to the ON position.
(2) Set the power switch to ON.
As this unit should be connected to the power outlet unit,
leave its power switch in the ON position.
Power unit
(1)
(2)
Содержание FLUOVIEW FV300
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 26: ......
Страница 28: ......
Страница 38: ......
Страница 114: ......
Страница 291: ...APPLIED OPERATIONS Protocol processor IV OPERATION INSTRUCTIONS IV 2 1 7 7 Page Fig 2 54 Panel After a Repeated Scan...
Страница 422: ...APPLIED OPERATIONS Viewing 3D Image IV OPERATION INSTRUCTIONS IV 2 3 0 8 Page Fig 2 133 Panel Showing Stereo 3D Images...
Страница 486: ......
Страница 506: ......
Страница 512: ......
Страница 518: ......
Страница 520: ......
Страница 548: ......
Страница 552: ......
Страница 554: ......
Страница 556: ......
Страница 562: ......
Страница 563: ......