Maintenance of Major System Units
/Scan Unit
V .
MAINTENANCE
V.
2 - 1
Page
2 Maintenanc e of Major System Units
This section describes the maintenance of the following units.
z
Scan unit
z
Laser scanning microscope
2-1 Scan Unit
The scan unit used with this system has been designed to allow replacement of the
BARRIER FILTER slider and DETECTION MODE slider in order to accommodate a
wide range of research. This section describes procedures for replacing these parts and
for adjusting the light axis after replacement.
2-1-1 Replacing the Barrier Filters Slider
To replace the BARRIER FILTERS slider, loosen the filter slider cover screws (10) and
remove the slider together with the slider cover.
Then replace the BARRIER FILTER slider, fit the filter cover, and tighten the screws.
Be sure to tighten the screws firmly, or noise may interfere with
images.
NOTE
(7) Cover screw for accessing
excitation cube unit
(Includes the cube locking
screw (8) inside.)
(2) DETECTION
MODE slider
(1) CONFOCAL APERTURE knob
(4) Excitation
DM selector
knob
(10) Filter slider cover
screws
(9) Cover screws for optical
axis adjustment
Содержание FLUOVIEW FV300
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 26: ......
Страница 28: ......
Страница 38: ......
Страница 114: ......
Страница 291: ...APPLIED OPERATIONS Protocol processor IV OPERATION INSTRUCTIONS IV 2 1 7 7 Page Fig 2 54 Panel After a Repeated Scan...
Страница 422: ...APPLIED OPERATIONS Viewing 3D Image IV OPERATION INSTRUCTIONS IV 2 3 0 8 Page Fig 2 133 Panel Showing Stereo 3D Images...
Страница 486: ......
Страница 506: ......
Страница 512: ......
Страница 518: ......
Страница 520: ......
Страница 548: ......
Страница 552: ......
Страница 554: ......
Страница 556: ......
Страница 562: ......
Страница 563: ......