Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
IV.
OPERATION INSTRUCTIONS
IV
. 1 - 2 4
Page
HeCd laser.
(1) Turn the rocker switch(1) of laser controller to “I” (ON).
(2) Turn the key switch(2) of laser controller to the ON position.
(3) Turn the key switch(3) of remote interface module to the ON
position.
(4) Turn the <ON> button of the remote interface module ON.
After the key is set to ON, it takes tens of seconds before the
laser oscillation starts.
LD405/440 laser.
(1) Confirm that the provided shorting plug is attached to the
remote interlock (1) or that the user’s equipment is
connected to the remote interlock and interlocking released.
(2) Set the power switch (2) to “|” (ON).
(3) Turn the Key (3) to the ON position.
(4) Set the shutter switch (4) to OPEN .
If the LASER EMISSION LED of the LD405/440
laser power supply is lit in red, the laser is
oscillating. In this case, simple setting the
shutter switch (4) to OPEN may cause the
laser beam to be output depending on
setups.
Laser controller
Remote interface module
LD405/440 laser power supply
(1)
(2)
(3)
(4)
NOTE
(2)
(4)
(1)
(3)
Содержание FLUOVIEW FV300
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 26: ......
Страница 28: ......
Страница 38: ......
Страница 114: ......
Страница 291: ...APPLIED OPERATIONS Protocol processor IV OPERATION INSTRUCTIONS IV 2 1 7 7 Page Fig 2 54 Panel After a Repeated Scan...
Страница 422: ...APPLIED OPERATIONS Viewing 3D Image IV OPERATION INSTRUCTIONS IV 2 3 0 8 Page Fig 2 133 Panel Showing Stereo 3D Images...
Страница 486: ......
Страница 506: ......
Страница 512: ......
Страница 518: ......
Страница 520: ......
Страница 548: ......
Страница 552: ......
Страница 554: ......
Страница 556: ......
Страница 562: ......
Страница 563: ......