Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
IV.
OPERATION INSTRUCTIONS
IV
. 1 - 4 6
Page
1-2-12-4 Setting the Highest Scan Speed
1. Set the scan speed to the fastest speed by using the scale in the [Scan Speed] group
box in the [Acquire] panel
The focus mode makes it possible to increase the line skipped scan speed.
From the page tabs on the bottom right of the [Acquire] panel, select the
[Settings] sub-panel.
Select either option button in the [Focus Mode] group box.
The focus mode is enabled when acquiring images using the <Focus>
button.
The Focus function reduces the scanning time by line skipped scan.
As a result, the acquired images become coarse.
[Scan Speed] group box
Set the scan speed by clicking a point on
the scale line.
NOTE
TIP
[Focus Mode] group box
[X2] option button
Acquires image at twice the
highest speed.
[X4] option button
Acquires image at 4 times the
highest speed.
Increasing the number of
divided images in the [Display]
panel, line skipped scan at 4
times (Focus) cannot be done.
Содержание FLUOVIEW FV300
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 12: ......
Страница 26: ......
Страница 28: ......
Страница 38: ......
Страница 114: ......
Страница 291: ...APPLIED OPERATIONS Protocol processor IV OPERATION INSTRUCTIONS IV 2 1 7 7 Page Fig 2 54 Panel After a Repeated Scan...
Страница 422: ...APPLIED OPERATIONS Viewing 3D Image IV OPERATION INSTRUCTIONS IV 2 3 0 8 Page Fig 2 133 Panel Showing Stereo 3D Images...
Страница 486: ......
Страница 506: ......
Страница 512: ......
Страница 518: ......
Страница 520: ......
Страница 548: ......
Страница 552: ......
Страница 554: ......
Страница 556: ......
Страница 562: ......
Страница 563: ......