OmegaPATPlus / BetaPATPlus Technical specifications
17
3.5 Touch
leakage
current
Touch leakage current readout
Range Resolution Accuracy
0.00 mA
÷
3.99 mA
0.01 mA
±
(10 % of r 5 digits)
Pass levels ............... 0.25 mA, 0.50 mA, 1.00 mA, 3.50 mA
Test duration............. 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, 180 s, none
Output....................... Class 1, Class 2: test socket – S test probe connector
Test method.............. Current measurement circuit according to VDE 0404-3, Appendix
A, Figure A1, IEC 60990 Figure F2
3.6 Polarity
test
Test voltage .............. < 50 V AC
Detects ..................... Pass, L-open, N-open, PE-open, L-N crossed, L-PE crossed, N-
PE crossed, L-N shorted, L-PE shorted, N-PE shorted, multiple
faults
Output....................... test socket – IEC socket
3.7 Clamp
current
True RMS current readout using 1000:1 current clamp
Range Resolution
Accuracy*
)
0.00 mA
÷
9.99 mA
0.01 mA
±
(5 % of r 10
digits)
10.0 mA
÷
99.9 mA
0.1 mA
±
(5 % of r 5 digits)
100 mA
÷
999 mA
1 mA
±
(5 % of r 5 digits)
1.00 A
÷
9.99 A
0.01 A
±
(5 % of r 5 digits)
10.0 A
÷
24.9 A
0.1 A
±
(5 % of r 5 digits)
*
)
Consider accuracy of current transformer.
Pass levels ............... 0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.25 mA,
2.50 mA, 3.00 mA, 3.50 mA, 5.00 mA, 9.90 mA, none
Test duration............. 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, 180 s
Input.......................... current clamp terminals