OmegaPATPlus / BetaPATPlus Measurements
33
6.4.1 How to perform single leakage current measurement
From the
Main menu
, select
Single test
and then select
Leakage
by using
¿
and
À
cursor keys and press ENTER key to confirm.
Leakage menu
is displayed.
Leakage menu
Measurement parameters are detailed in the upper right corner of the display. The
following parameters can be adjusted for this measurement:
Ì
high limit leakage value,
Ì
measurement
duration.
To edit parameters press F1 function key first. Use
¿
and
À
cursor keys to select the
parameter you want to edit. By using
½
and
¾
cursor keys adjust the selected
parameter. Press F1 function key again to confirm the selected settings.
Press the START key to perform a measurement.
The measurement can be aborted by pressing STOP key at any time. Actual result is
shown on the display during measurement. After measurement time period has elapsed
the last measured value is displayed, and PASS or FAIL indication appears on the
display (based on the high limit leakage value).
Examples of leakage current and apparent power measurement results
Press START key to repeat the test or press ESC key to return to the
Single test
menu
.
Notes:
Ì
During the test, a mains voltage is connected to the appliance. If appliance
contains moving parts, make sure that it is safely mounted or protected to
prevent possible danger to the operator or damage to the appliance or
surrounding environment!
Ì
Consider any displayed warning before starting measurement!