6.9 Flash test (MI 3305 – OmegaPATPlus only)
The purpose of this test is to ensure that the insulation material of the appliance can
withstand high voltage transients and short overvoltages.
6.9.1 Flash test on class I appliances
For Class I appliances, a test voltage of 1.5 kV
AC
is applied between main test socket
(L+N) and PE test terminals.
Flash test on Class I appliance
6.9.2 Flash test of class II appliances
For Class II appliances, a test voltage of 3 kV
AC
is applied between main test socket
(L+N) and FLASH test terminals.
Flash test on Class II appliance
6.9.3 How to perform single Flash measurement
From the
Main menu
, select
Single test
and then select
Flash test
by using
¿
and
À
cursor keys and press ENTER key to confirm.
Flash menu
is displayed.