OmegaPATPlus / BetaPATPlus
Autotest sequences
56
7.4.3 Insulation resistance measurement
Notes:
Ì
For class II appliances the test probe should be used (see chapter
6.2.2
Insulation resistance on class II appliances
)!
Ì
Consider the displayed warnings before starting measurement!
If the insulation resistance measurement and all previous tests passes (except visual
test), the instrument will automatically proceed onto the next test (in single
measurement mode only).
If the insulation measurement fails or the test is set to continuous measurement mode,
the autotest sequence stops and a prompt will appear on the screen (in continuous
measurement mode only).
The following actions can be taken to continue or complete the autotest sequence:
START...................... Start the insulation resistance measurement.
Proceed with the next insulation resistance measurement (in
continuous measurement mode only).
Proceed to the next autotest sequence measurement (in single
measurement mode only).
F2 function key...........View insulation resistance test help screens.
F3 function key ......... Repeat the insulation resistance measurement.
ENTER key............... Proceed to the next autotest sequence measurement (in
continuous measurement mode only).
F4 function key ......... Skip insulation resistance measurement.
F5 function key ......... End the autotest sequence.
7.4.4 Substitute leakage current measurement
Notes:
Ì
For class II appliances the test probe should be used! (see chapter 6.3.2
Substitute leakage current on class II appliances
)
Ì
Consider the warnings on the display before starting measurement!
The autotest sequence stops when substitute leakage current measurement is
performed.
The following actions can be taken to continue or complete autotest sequence:
START...................... Start the substitute leakage current measurement.
Proceed with the next substitute leakage current measurement (in
continuous measurement mode only).
Proceed to the next autotest sequence measurement (in single
measurement mode only).
F2 function key...........View substitute leakage test help screens.
F3 function key ......... Repeat the substitute leakage current measurement.
ENTER key............... Proceed to the next autotest sequence measurement (in
continuous measurement mode only).
F4 function key ......... Skip substitute leakage current measurement.
F5 function key ......... End the autotest sequence.