OmegaPATPlus / BetaPATPlus Measurements
35
Measurement parameters are detailed in the top right corner of the display. The
following parameters can be adjusted in this measurement:
Ì
high limit leakage value,
Ì
measurement
duration.
To edit parameters press F1 function key first. Use
¿
and
À
cursor keys to select the
parameter you want to edit. By using
½
and
¾
cursor keys adjust the selected
parameter. Press F1 function key again to confirm the selected settings.
Press the START key to perform a measurement.
The measurement can be aborted by pressing STOP key at any time. The actual result
is shown on the display during the measurement. After the measurement time period
has elapsed the last measured value is displayed, and PASS or FAIL indication appears
on the display (based on the high limit leakage value).
Examples of touch leakage current measurement results
Press START key to repeat the test or press ESC key to return to the
Single test
menu
.
Notes:
Ì
During the test, a mains voltage is connected to the appliance. If appliance
contains moving parts, make sure that it is safely mounted or protected to
prevent possible danger to the operator or damage to the appliance or
surrounding environment!
Ì
Consider any displayed warning before starting measurement!
6.6 Polarity
test
This test checks the polarity of a mains supply cable. The following faults can be
detected: L open, N open, PE open, L-N crossed, L-PE crossed, N-PE crossed, L-N
shorted, L-PE shorted, N-PE shorted, multiple faults.