OmegaPATPlus / BetaPATPlus
Autotest sequences
50
7.2.2 How to save the autotest custom sequence
In the
Autotest save menu
, the following actions can be taken:
Ì
an existing autotest sequence can be saved or renamed,
Ì
a new autotest sequence can be added to the list.
In
Autotest custom view menu
press F2 function key to save or rename the edited
autotest sequence or press F3 function key to add a new autotest sequence to the list.
The
Autotest save menu
is displayed.
Autotest custom save menus
When saving or renaming an autotest sequence, the existing name of the selected
custom procedure is shown in
Name sets
field. Up to 17 alpha-numeric and special
characters can be entered in
Name sets
field. Press ESC key to return
Autotest
custom menu
.
After the autotest name has been entered, press the F1 function key for confirm or the
ESC key to return to the
Autotest custom menu
. To restore last used name press F2
function key.
7.2.3 How to delete an existing custom test sequence from the list
In
Autotest custom menu
, select the test sequence you want to delete by using
¿
and
À
cursor keys. Press the F3 function key to delete. Confirm the delete by pressing the Y
key. The selected autotest sequence will then be deleted from the list.
Notes:
Ì
The first custom autotest sequence cannot be deleted from the list.
Ì
If only one custom autotest is defined, it cannot be deleted from the list!
7.2.4 How to start custom autotest sequence
An autotest can be started from the
Autotest custom menu
by following a simple
procedure:
In
Autotest custom menu
, select the test sequence to be executed (see
7.1.1
Selecting the autotest shortcut sequence
). Press the START key to start the autotest
sequence. Refer to chapter
7.4 Performing autotest sequences
for more information.