OmegaPATPlus / BetaPATPlus Measurements
45
Flash test menu
The measurement parameters are detailed in the top right corner of the display. The
following parameters can be adjusted in this measurement:
Ì
output voltage value,
Ì
current limit value,
Ì
measurement
duration.
To edit parameters, press F1 function key first. Use
¿
and
À
cursor keys to select the
parameter you want to edit. By using
½
and
¾
cursor keys adjust the selected
parameter. Press F1 function key again to confirm the selected settings.
Press the START key to perform a measurement.
The measurement can be aborted by pressing STOP key at any time. The actual result
is shown on the display during the measurement. In case, that current limit is exceeded
during the test, measurement is aborted immediately. After the measurement time has
elapsed or aborted the last measured value is displayed, and PASS or FAIL indication
appears on the display (based on the current limit leakage value).
Examples of flash test results
Press START key to repeat the test or press ESC key to return to the
Single test
menu
.
Notes:
Ì
Consider any warning on the display before starting measurement.
Ì
Do not touch the equipment under test or the test leads during the test – a
dangerous voltage will be present.
Ì
Appliance
ON/OFF
switches must be switched on.
Ì
Before applying the test, check that the components in the appliance can
withstand the test voltage without any damage. IT appliances are especially
sensitive.
.