OmegaPATPlus / BetaPATPlus
Autotest sequences
53
While recalling stored results, the instrument shows a bargraph and a ratio of files found
compared to files stored in memory (e.g. 7/11 implies 7 results have been found to meet
the filter criteria out of a potential 11 results stored in the flash memory).
By using
¿
and
À
cursor keys, the F1 and F2 function keys or barcode reader, select
the appliance that should be retested.
Appliance tickers
Each appliance is marked with a ticker. The ticker appears at the right of the appliance
number and helps to speed up finding appliances to be retested.
The meaning of the tickers is as follows:
PC
The stored appliance data was uploaded from PC
PC
9
The stored appliance data was uploaded from PC and has been retested
I
The stored appliance data was performed with the instrument and stored.
I
9
The stored appliance data was performed with the instrument and stored and
has been retested.
By pressing ENTER key on an appliance, more information regarding the appliance
results and data can be viewed. The information can be scrolled using the F1 and F2
function keys and the F5 function key can be used to switch between different appliance
data views.
Aus/NZ model
UK model
View results menu examples
Use ESC key to return to previous menu.
7.3.2 How to start a project autotest
From the
Main menu
, select
Project autotest
. Using the filters, search for the
appliances to be retested. To retest a particular appliance, find the appliance by using
the
¿
and
À
cursor keys, the F1 and F2 function keys or the barcode reader and begin
retesting the appliance by pressing the START key.
If the appliance is new, a standard autotest can be performed as decribed in
7.4
Performing autotest sequences
:
Ì
the autotest sequence will be the same as it was for the selected Project
autotest.
Ì
exactly the same appliance data (except time and user) will be offered to store in
the selected Project autotest.
Ì
when a new autotest is saved, it will get an »
I
« ticker. The original autotest will
get a »
I
9
« ticker when the appliance is retested through the project autotest.