OmegaPATPlus / BetaPATPlus Instrument description
12
Leakage test probe-PE
too high!
An exceptionally high leakage current (higher than about 5
mA) was measured through S test terminal during the pre-
test.
External voltage on test
socket too high!
DANGER!
Voltage on test socket or LN/PE terminals is higher than
20 V
(AC or DC) approximately!
Disconnect the appliance under test from the instrument
immediately and determine why external voltage was
detected!
External voltage on test
probe too high!
DANGER!
Voltage on test probe (S) is higher than 25 V
(AC or DC)
approximately!
Disconnect the test probe from the appliance and
determine why external voltage was detected!
Test was skipped for
safety!
Instrument skipped the required test because of a failed
previous test.
Instrument overheated!
Measurement is
aborted!
Temperature of internal components of the instrument
reached their top limit. Measurement is prohibited until the
internal temperature has reduced.