OmegaPATPlus / BetaPATPlus Single test mode
25
5 Single test mode
In single test mode, individual tests can be performed. This is especially helpful when
the test engineer suspects that one or more faults are present in the appliance under
test.
Note:
Ì
Test results cannot be saved in this mode!
5.1 How to perform measurements in single test mode
Select
Single test
in
Main menu
by using
¿
and
À
cursor keys and press ENTER key
to confirm the selection. The
Single test menu
is displayed.
Single test menu
By using
¿
and
À
cursor keys select the measurement you want to perform and press
ENTER key to confirm the selection. To return to the
Single test menu
press ESC key.
Selected test measurement parameters are detailed in the top right corner of the
display. These parameters can be edited by pressing the F1 button. In general the
following parameters can be set:
Ì
output
test
voltage/current,
Ì
pass
level,
Ì
measurement
duration.
Once the test parameters have been set, the settings are saved until new changes are
made or the instrument is reset to the default configuration.
Note:
Ì
To keep new settings in the case of a power failure, return to main menu or press
SAVE before turning the power off.