OmegaPATPlus / BetaPATPlus Technical specifications
18
3.8 Trip-out time of portable RCD
Portable RCD trip-out time readout
Range Resolution
Accuracy
0 ms
÷
1999 ms (½
×
I
Δ
N
)
1 ms
0 ms
÷
300 ms (I
Δ
N
)
1 ms
0 ms
÷
40 ms (5
×
I
Δ
N
)
1 ms
±
3 ms
Test currents (I
Δ
N
) ............. 10 mA, 15 mA, 30 mA
Test current multipliers ..... ½
×
I
Δ
N
, 1
×
I
Δ
N
, 5
×
I
Δ
N
Test modes....................... single, autotest
Output............................... PRCD probe connector – test socket
Trip out times according to EN 61008 / EN 61009 (Aus/NZ models)
½
×
I
Δ
N
*)
I
Δ
N
5
×
I
Δ
N
t
Δ
> 300 ms
t
Δ
< 200 ms
t
Δ
< 40 ms
Trip out times according BS 7671 (UK models)
½
×
I
Δ
N
*)
I
Δ
N
5
×
I
Δ
N
t
Δ
> 1999 ms
t
Δ
< 200 ms
t
Δ
< 40 ms
*)
The ½
×
I
Δ
N
current does not trip out the RCD.
3.9 Flash test (MI 3305 – OmegaPATPlus only)
Flash current readout
Range Resolution Accuracy
0.00 - 2.50 mA
0.01 mA
±
(5 % of r 5 digit)
Test voltage ...................... 1500 V AC, 3000 V AC
Output resistance.............. 480 k
Ω
@1500 V, 960 k
Ω
@3000 V
Limits ................................ 1.00 mA, 1.50 mA, 2.00 mA, 2.25 mA
Timer ................................ 2 s, 3 s, 5 s, 10 s, 30 s
Output............................... Class 1: test socket
Class 2: test socket (L+N) – flash probe connector
3.10 General data
Rated supply voltage .........................230 V
AC
/ 115 V
AC
(
±
10 %, 50 or 60 Hz)
Max. power consumption.................. 150 VA (earth bond 25 A short-circuit, without
tested appliance)
Maximum appliance current...............16 A
Overvoltage category.........................Cat II / 300 V
Protection classification .....................I
Pollution degree.................................2