OmegaPATPlus / BetaPATPlus Autotest shortcut parameters
83
FUSE:
3 A (750 VA), 6 A (1.5 kVA), 10 A (2.5 kVA), 13 A (3.2 kVA), unknown
SUPPLY CORD:
short, middle (or low c.s.a.), long
Cord
Earth bond
limit
Appliance
fuse
Power
Nominal
conuctor
c.s.a.
Appliance
supply cord
length
3 A
750 VA
0.5 mm
2
(39 m
Ω
/m)
2.5 m
6 A
1.5 kVA
0.75 mm
2
3.5
m
10 A
2.5 kVA
1 mm
2
5
m
SHORT
0.10
Ω
13 A
3.2 kVA
1.25 mm
2
6
m
3 A
750 VA
0.5 mm
2
7.5
m
6 A
1.5 kVA
0.75 mm
2
11.5
m
10 A
2.5 kVA
1 mm
2
15
m
MIDDLE
0.30
Ω
13 A
3.2 kVA
1.25 mm
2
19
m
3 A
750 VA
0.5 mm
2
12.5
m
6 A
1.5 kVA
0.75 mm
2
19
m
10 A
2.5 kVA
1 mm
2
25
m
LONG
0.50
Ω
13 A
3.2 kVA
1.25 mm
2
32
m
Nominal conductor c.s.a. Nominal
conductor
resistance
0.5 mm
2
39 m
Ω
/m
0.75 mm
2
26 m
Ω
/m
1 mm
2
19.5 m
Ω
/m
1.25 mm
2
15.6 m
Ω
/m
See
Code of practice for in-service inspection and testing of electrical equipment, 3rd
edition
, page 124
Ì
The earth bond limit depends signlificantly on the supply cord length and its PE
wire cross-section area.
Ì
SHORT should be set for appliances with a cord not longer than 5 m.
Ì
MEDIUM can be used when the appliance supply cord is longer than 5 m or the
cord wire has a small cross-section area.
Ì
LONG can be used if the appliance has a very long supply cord (for instance
vaccum cleaners).
Ì
The power means the maximum power that can be delivered in combination with
the set fuse. Some users are more familiar with the nominal power than the fuse
rating.
FUNCTIONAL / PROTECTIVE CONDUCTOR TEST:
yes, no
LENGTH:
≤
5 m, 7.5 m, 10 m, 12 m, 15 m, 20 m, 30 m, 40 m, 50 m
CONDUCTOR CROSS-SECTION:
0.5 mm
2
, 0.75 mm
2
, 1.0 mm
2
, 1.25 mm
2
, 1.5 mm
2
,
unknown
MAX. CURRENT CAPACITY:
3 A (<750 VA), 6A (1.5 kVA), 10 A (2.5 kVA), 13 A (3.2
kVA), 15 A (3.75 kVA)