OmegaPATPlus / BetaPATPlus Technical specifications
16
3.3 Substitute leakage current
Substitute leakage current readout
Range Resolution
Accuracy
0.00 mA
÷
19.99 mA
0.01 mA
±
(5 % of r 5 digits)
Open circuit voltage .. < 50 V AC at mains voltage of 230 V AC
Short circuit current... < 40 mA
Pass levels ............... 0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.25 mA, 3.50
mA, 7.00 mA, 9.90 mA, 15.00 mA
Test duration............. 2 s, 3 s, 5 s, 10 s, 30 s
Output....................... Class 1: test socket
Class 2: test socket (L+N) – S test probe connector
Test method.............. floating to earth
Displayed current is calculated to appliance mains supply voltage.
3.4 Leakage current and apparent power
Differential leakage current readout
Range Resolution
Accuracy
0.00 mA
÷
9.99 mA
0.01 mA
±
(5 % of r 5 digits)
Pass levels ............... 0.25 mA, 0.50 mA, 0.75 mA, 1.00 mA, 1.50 mA, 2.25 mA,
2.50 mA, 3.00 mA, 3.50 mA, 5.00 mA, 9.90 mA
Test duration............. 2 s, 3 s, 5 s, 10 s, 30 s, 60 s, 120 s, 180 s, none
Output....................... Class 1: test socket
Test method.............. Current measurement circuit according to VDE 0404-3, Appendix
A, Figure A1, IEC 60990 Figure F2
Apparent power readout
Range Resolution Accuracy
0.00 kVA
÷
4.00 kVA
0.01 kVA
±
(5 % of r 3 digits)
Output....................... test socket