10 Maintenance
10.1 Periodic calibration
It is essential that all measuring instruments are regularly calibrated in order for the
technical specification listed in this manual to be guaranteed. We recommend an annual
calibration. The calibration should be done by an authorized technical person only.
10.2 Fuses
There are two fuses available from front panel:
F1 = F2 = T 16 A / 250 V (32
×
6,3 mm): intended for instrument protection.
If the instrument does not respond after connection to mains supply, disconnect the
mains supply and accessories and then check these fuses. For position of fuses refer to
chapter
2.1 Front panel
.
Warning!
Ì
Switch off the instrument and disconnect all test accessories and mains cord
before replacing the fuses or opening the instrument. Disconnect all test leads
and the instrument mains cord before removing fuse cover!
Ì
Replace blown fuse with the same type.
10.3 Service
For repairs under or out of warranty time please contact your distributor for further
information.
Unauthorized person is not allowed to open the OmegaPATPlus / BetaPATPlus
instrument. There are no user replaceable components inside the instrument.
10.4 Cleaning
Use a soft, slightly moistened cloth with soap water or alcohol to clean the surface of
OmegaPATPlus / BetaPATPlus instrument. Leave the instrument to dry totally before
using it.
Notes:
Ì
Do not use liquids based on petrol or hydrocarbons!
Ì
Do not spill cleaning liquid over the instrument!