OmegaPATPlus / BetaPATPlus
Autotest sequences
46
7 Autotest sequences
Autotest is the fastest and easiest way to test appliances. A pre-programmed autotest
sequence runs all the tests needed for verifying that the selected appliance is safe to
use. The complete autotest results can be stored together with their associated
appliance name and all related information.
7.1 Autotest shortcut menu
In this autotest menu any of the pre-defined autotest procedures can be selected. Pre-
programmed sequences cover almost all
in-service
tests, regardless of appliance type
and safety class.
All limits and tests are in compliance with currently valid standards and regulations. In
case of any changes, new firmware will be available at your distributor or from Metrel
directly.
The Autotest shortcut sequences depend on the OmegaPATPlus / BetaPATPlus model.
The complete list of autotest shortcut sequences can be found in Appendices A and B
of this manual.
7.1.1 Selecting the autotest shortcut sequence
From the
Main menu
select
Autotest shortcut
by using
¿
and
À
cursor keys, and
press ENTER key to confirm.
Autotest shortcut menu
is displayed.
Aus/NZ model
UK model
Autotest shortcut menu examples
The autotest sequence can simply be selected with three-digit test code using
½
and
¾
cursor keys or entering three-digit code by numeric keypad.
If the appropriate test code is unknown, the appliance sequence can be selected based
on the appliance type and safety class. To set the autotest sequence code or appliance
parameters, first use
¿
and
À
cursor keys to select the parameter you want to edit. The
½
and
¾
cursor keys can then be used to adjust the selected parameter.
The three-digit autotest code can also be selected by using the barcode reader. In this
case connect barcode reader to the DB-9 female BARCODE connector first.