OmegaPATPlus / BetaPATPlus
Autotest sequences
47
Connecting barcode reader to the BetaPAT instrument
The code read from the barcode label is accepted when two short beeps are emitted.
All test parameters and limits can simply be viewed on the display. For more information
refer to chapter
7.1.2 How to view autotest shortcut sequence measurement
parameters
.
7.1.2 How to view autotest shortcut sequence measurement
parameters
After pressing the F1 function key in
Autotest shortcut menu
the measurement
parameters of the selected autotest sequence are detailed.
Aus/NZ model
UK model
Autotest shortcut view menu
By using
¿
and
À
cursor keys, the parameters of the selected measurement can be
viewed.
The following measurement parameters can be viewed in all measurements:
Ì
measurement
mode,
Ì
output test voltage or current (except in visual test and TRMS current
measurement),
Ì
pass level (except in visual test),
Ì
measurement duration (except in visual test).
When
mode
is set to
disable
, the selected measurement is skipped.
When
mode
is set to
single
, one measurement will be performed during the
autosequence.
When
mode
is set to
continuous
, up to five repetitive measurements can be performed
(except earth bond measurement where up to six).