OmegaPATPlus / BetaPATPlus Other instrument operations
79
By using ESC key in the
Communication port
or
Baud rate submenu
, the instrument
will return to
Communication menu
without any changes.
Note:
Ì
Only one port can be active at one time.
9.3.9 Reset instrument settings
In this menu the following parameters can be set to their initial values:
Ì
all measurement parameters in single test mode,
Ì
user defined tests are cleared,
Ì
8 autotest sequences with all measurements disabled are added,
Ì
all menus are set to their initial structure,
Ì
PC baud rate is set to 115000 bps,
Ì
printer protocol is set to hardware (DTR).
Select
Original settings
in
Setup menu
by using
¿
and
À
cursor keys and press
ENTER key to confirm. The following menu is displayed:
Original settings menu
Select YES to confirm, or NO to quit without changes.