OmegaPATPlus / BetaPATPlus
Autotest sequences
49
Aus/NZ model
UK model
Autotest custom view menu
Depending on the measurement the following measurement parameters can be set
here:
Ì
measurement
mode,
Ì
output test voltage or current (except in visual test and TRMS current
measurement),
Ì
pass level (except in visual test),
Ì
measurement duration (except in visual test).
When
mode
is set to
disable
, the selected measurement is skipped.
When
mode
is set to
single
, one measurement will be performed during the
autosequence.
When
mode
is set to
continuous
, up to five repetitive measurements can be performed
(except earth bond measurement where up to six).
By using
¿
and
À
cursor keys in
Autotest custom view menu
select the
measurement you want to edit. Press the F1 function key to enable parameters editing
of selected measurement. Use
¿
and
À
cursor keys to select parameter you want to
edit. Then use
½
and
¾
cursor keys to adjust selected parameter.
Adjusting measurement parameters
Press F1 function key to accept the new settings for selected measurement. The
instrument will then return to
Autotest custom view menu
.
Once the parameters for all the required measurements have been set, existing autotest
sequence can be saved under the same name or renamed. Also new autotest
sequence can be added to the list. Refer to chapter
7.2.2 How to save the autotest
custom sequence
.