34
Nexam Pro User Manual
Chapter 2
Figure 2.13
Transparent view of pump unit
1. Lever to press the tube clamp over the tube
2. Mark on tube clamp to indicate 5-roller or
10-roller pump
3. Tube clamp
4.
Rollers
5.
Control panel
6.
Flush button
7.
Weighing hook (infused volume transducer)
Install the pump tube as follows:
▪
Start the pre-test of the investigation.
▪
Turn the lever of the pump [1] and take out the tube clamp [3].
▪
Unpack the pump tube.
▪
Put the spike of the pump tube into the bag of saline.
▪
Hang the bag of saline on the hook [7] of the pump unit.
▪
Put the pump tube on the rollers; the tube has a special part which must be
put on the rollers, indicated by two stoppers.
Figure 2.14
Nexam Pro pump tube with two stoppers
Stopper