Biofeedback
Nexam Pro User Manual
165
Figure 10.3
Pretest window Biofeedback on EMG
For EMG continue as follows:
❑
Apply EMG electrodes / insert probe.
❑
Ask the patient to squeeze the muscles to check the EMG signal.
For pressure continue as follows:
❑
Prepare the catheter / probe.
Zero the transducer in open air.
❑
Insert the catheter / probe.
❑
Ask the patient to cough to check the registration of the pressure.
Auto scale
It is possible to set the scale automatically as follows:
❑
Let the patient relax for a while.
❑
Ask the patient to squeeze, for example the pelvic floor / anal
sphincter, as much as possible for a while.
Click the
Autoscale
button. The minimum and maximum values, which
are measured over the previous 10 seconds, will be used to set the
Minimum scale
Maximum scale
Actual measured value
Image representing
the current value
General and More tab
Heart beat filter