Urethral Pressure Profile (UPP)
Nexam Pro User Manual
99
7.4.2
Stress Profile
For each stress profile, the following parameters are default calculated:
•
Average transmission
•
Resting bladder pressure
•
Maximum urethral closure pressure (MUCP)
•
Maximum urethral pressure (MUP)
•
Functional profile length
Figure 7.2
UPP results selection
The parameters of the UPP results are described below.
Maximal urethral pressure (MUP)
The maximal urethral pressure (P
ura
) during the UPP study.
Functional profile length
The functional profile length is the length of the urethra, along which the
urethral pressure exceeds vesical pressure. This urethral length is measured
in millimeters. It is calculated from the puller speed in conjunction with the time
needed to reach the point with the same pressure as the resting bladder