36
Nexam Pro User Manual
Chapter 2
NOTE
For correct functioning of the pump, it is essential that the
bag of saline
hangs freely
on the hook of the pump.
When the pump is running, the bag should not be
touched, because this can affect the infused volume
measurement.
NOTE
Ensure that the correct specific weight of the infusion
liquid is set in the investigation protocol. This is essential
as the infused volume is calculated by determining the
decrease in weight of the infusion liquid. If a video-mix
option is fitted, the system will automatically calculate
infused volume using the specific weight entered for the
contrast liquid if, and only if, the video-mix option is
switched ON.
Figure 2.16
Install the pump tube
Pump stop
Under normal situations, the pump stops automatically when the bladder
pressure Pves is over a limit >2 seconds. This limit (Max. Pves) can be set in
Содержание Nexam Pro CIM
Страница 1: ...Nexam Pro User Manual...