Safety Information
Nexam Pro User Manual
249
CAUTION
Remove the batteries when the flowsensor is not used for a
longer period (more than three months).
CAUTION
Do not use chloride-based solutions for cleaning the flow
stand. Check that the solution does not affect stainless
steel (AISI304L). Clean the flow stand if it was
contaminated with urine.
NOTE
Environmental conditions (i.e., thick walls/doors, iron in
walls/doors) may compromise the Bluetooth connection of
the flowsensor and battery lifetime.
Wireless (Bluetooth) puller (Mk III)
The Bluetooth puller complies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation. This device
contains FCC-ID POOWML-C40XX.
WARNING
Make sure that the Bluetooth puller is connected to the
correct power adapter (PS-BT-PULLER). Do not use a
power adapter other than that which is specified. Doing
so may cause fire leading to serious damage.
WARNING
When adjusting the spring balanced arm be aware not to
pinch fingers or other objects.
WARNING
Connect the power adapter whenever the Bluetooth
puller is not used to prevent batteries to become
depleted. Deep discharging the batteries will damage the
batteries internally and shorten the lifespan.
WARNING
If the slider does not move smoothly (ticking noises)
during an investigation, the results are not reliable. Solve
the problem and redo the investigation.