Anorectal Manometry
Nexam Pro User Manual
143
Press the
ALT
key on the remote control and press the
Calibration
begin
button. Inflate the balloon with 10 ml increments, until you reach
the maximum volume which you want to inflate the balloon during the
investigation.
Press the
Calibration end
button and empty the balloon.
Press the
Stop investigation
button.
The balloon compliance will be
saved and is used by the next investigation.
During analysis the software will compensate the rectal pressure with the
pressure built up by the balloon. If you always use the same balloon catheter,
you need to determine the balloon compliance only once.
Measurement
The balloon should now be inflated with 10 ml increments and the
resulting pressure at each step is recorded.
If you want to increment in other steps than 10 ml, press the
pump
up/down keys on the remote control. The inflation step is displayed on
the syringe button on top of the screen.
After inflation of the balloon, a 20-second accommodation period
should be provided.
During infusion the sensation markers can be set by pressing the
appropriate softkey. When the maximum tolerable volume is reached,
the investigation can be stopped and the balloon can be deflated.
Filling with LABORIE Perfusion pump plus (when available)
•
On the Filling tab:
Filling method: Perfusion pump (Automatic, Air)
•
On the Sensation/Compliance tab:
Use std. Balloon calibration
Always
Save std. Balloon calibration
Always
Содержание Nexam Pro CIM
Страница 1: ...Nexam Pro User Manual...