122
Nexam Pro User Manual
Chapter 9
Figure 9.11
Sensation settings
During the sensation test different sensation markers can be placed. The
following markers can be enabled in the protocol: no sensation, first sensation,
first urge, modest urge, intense urge, maximum tolerable volume, and pain.
The setting
display balloon markers
can be (de)selected.
Balloon filling settings
On the
Filling
tab, you can set the filling method. For the syringe filling
method, you can enter the volume increment (e.g., 10 ml) or edit a complete
list (e.g., 10-10-10-20-20) on the
Filling
tab. You can also select
Ask balloon
volume
. After inserting a sensation marker, during the test the software will
ask to enter the balloon volume.
For filling the balloon with the pump (water) you can enter the pump speed
(default 50 ml/min) on the
Filling
tab.
9.2.10 Compliance
Click the
More
button and then select the
Compliance
tab. The window
shown in Figure 9.12 is displayed.