52
Nexam Pro User Manual
Chapter 2
The control panel of the wireless puller contains three LED’s: for Bluetooth [1],
Status [2] and Battery [3] indication, and one membrane button [4] to unlock
the slider.
1.
Bluetooth LED
2.
Status LED
3.
Battery LED
4.
Unlock slider button
Figure 2.40
Control panel
LEDs and button wireless puller
The following table describes the LEDs.
LED
Color
Indication
Bluetooth
Blue blinking
The Bluetooth catheter puller has a wireless
connection with the computer.
Status
Green on
The puller is switched on and is functioning
well.
Orange on
IPL mode and/or downloading firmware.
Red on
The puller has detected an internal failure.
Contact your LABORIE representative.
Battery
Green on
The rechargeable batteries are fully charged.
Green blinking
The power adapter is connected to the puller
(and charging the batteries).
Orange blinking The batteries are almost empty; connect the
power adapter (within five minutes).
Содержание Nexam Pro CIM
Страница 1: ...Nexam Pro User Manual...