SERVICE MANUAL
4-68
9G9002 – 2011-08-15
Condition Code: 27-651 – Exposure timeout notification
Condition Code: 28-501 – Processor RTD Short Circuit
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
None
Image Exposure timed
out, job will be reprinted
NORMAL
3-Notify
Description
This notification occurs when the image exposure does not complete in the
required time. The faulty image prints. The job fails and the imager attempts to
print the image again.
If this problem is detected on three successive films, error 27-650 is declared.
System Impact
None. The system will function normally. [SI-0]
User Action
None
Show at DS
False
DICOM Printer
Status
NORMAL
Unique Symb. ID
ERR_EXPOSE_TMO_RETRY
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
Processor: Internal
Hardware Failure
Short in RTD circuit
PROC DOWN
0-Fatal
Description
This error can occur for many electromechanical reasons: Resistance
Temperature Detector (RTD) and connector leads touching, drum-slip-ring
failure, condensation (cold machine in warm environment), failed component in
analog-to-digital conversion circuitry on the processor board, etc.
The error clears only when imager power is cycled.
System Impact
The imager can only print jobs that were in process when the error occurred. The
spooler can still accept jobs, but the jobs will not print until the error is resolved.
All jobs that were in process when the error occurred will reprint completely
when the error is resolved. [SI-1]
User Action
1. Shut down the laser imager, and then restart the imager.
2. If the error persists, call for service.
Show at DS
True
DICOM Printer
Status
FAILURE
Unique Symb. ID
ERR_PROCESSOR_RTD
Logged for RMS
[TriggerValues=a, b][TriggerNotes=RTD zone number, temperature in
degrees C]