Adjustments and Replacements
9G9002 – 2011-08-15
2-43
Power Supply Board
Removal
Possible damage from electrostatic discharge.
ESD
When installing circuit boards, always use a wrist strap to prevent damage from electrostatic
discharge.
[1]
Remove the:
• 2 power supply shield screws
• Power supply shield
[2]
Disconnect the input and output connectors, 1 on each side of the power supply board.
[3]
Remove the:
• 4 screws that secure the power supply board to the standoffs
• Power supply board
Installation
Reverse the steps in the removal procedure.
Prerequisites:
• Close the film cartridge. (Touch the Pause button on the display screen.)
• De-energize the laser imager.
• Disconnect the power cord.
•
Remove the back cover
.
Tools:
T10 Torx driver
H221_6021BC
Shield Screw
(1 of 2)
Power Supply Shield
Connectors
Power Supply
Screw (1 of 4)
Power Supply
Board