Adjustments and Replacements
9G9002 – 2011-08-15
2-41
Exposure Transport Belt
Removal
[1]
Remove the:
(a)
5 cover screws
(b)
Exposure transport cover
[2]
Push down on the tensioner to release pressure on the belt.
[3]
Slide the belt off the lower shaft and remove the belt.
Installation
Reverse the steps in the removal procedure.
Prerequisites:
Remove the:
•
Top cover
•
Right cover
Tools:
T10 Torx driver
Cover
Screws
Cover
Screws
Tensioner
Shaft
Belt