Diagnostics
9G9002 – 2011-08-15
4-19
Condition Code: 04-302 – Invalid Attribute
Condition Code: 05-000 – Error Unknown
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
None
Invalid attribute xxxx
(GGGG,EEEE)
encountered in YYYY
SOP (where xxxx
(GGGG,EEEE) is the
tag name and number,
and YYYY is SOP
name)
NORMAL
1-Alert
Description
An invalid attribute for the specified SOP class was found. For a complete
description of valid attributes and their contexts, see [PART4].
System Impact
The named SOP will not be processed. If the SOP is a film session or film box,
the requested page(s) will not print.
User Action
None.
Show at DS
False
DICOM Printer
Status
NORMAL
Unique Symb. ID
ERR_INVALID_ATTRIBUTE
Note
Rationale for not reporting to user: This is a DICOM SCU configuration problem,
not an operational error. The user cannot correct this from the imager. (This
problem likely occurs only on first-time SCU setup.)
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
None
Error Unknown
NORMAL
2-Warning
Description
An unknown error occurred in the Study Reclamation subsystem.
System Impact
Unknown. This is an unexpected or unknown error; it is not possible to
document specific details.
User Action
None
Show at DS
False
DICOM Printer
Status
NORMAL
Unique Symb. ID
ERR_UNKNOWN