SERVICE MANUAL
3-4
9G9002 – 2011-08-15
Performing the PM
Every PM, perform all of the following procedures in the order listed:
[1]
Print a flat-field gray film to compare against the film after PM is complete.
(a)
In the service tool, select
Diagnostics > DRE > Advanced Test Print
.
(b)
In the Test Profile drop-down list, select
Flat1.8
.
(c)
Click
Run
.
[2]
Touch the Pause icon on the display screen to close the film cartridge.
[3]
Open the film supply and remove the film cartridge. Leave the film supply open.
[4]
De-energize the imager, and then disconnect the power cord.
[5]
Remove the:
• Top cover (see Adjustments and Replacements,
“Top Cover”
).
• Left cover (see Adjustments and Replacements,
“Left cover”
).
• Right cover (see Adjustments and Replacements,
“Right Cover”
).
Caution
The processor drum is very hot. Allow it to cool for 10 minutes before cleaning.
[6]
Pivot open the processor drum cover
.
Continue with the next steps while the processor drum cools.
[7]
Remove the deodorant filter.
[8]
Use a vacuum cleaner to clean the:
• Deodorant filter area
• Filter duct
• Exhaust plenum