SERVICE MANUAL
4-70
9G9002 – 2011-08-15
Condition Code: 28-510 – Processor RTD Open Circuit
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
Processor: Internal
Hardware Failure
Processor detected a
sudden change to
infinite resistance in the
RTD circuitry
PROC DOWN
0-Fatal
Description
This error likely occurred in conjunction with the overheat error. The most likely
cause is an opening in the RTD circuit (RTD fractured, drum slip ring failure,
wiring harness disconnected, etc.).
The error clears only when imager power is cycled.
System Impact
The imager can only print jobs that were in process when the error occurred.
The spooler can still accept jobs, but the jobs will not print until the error is
resolved. All jobs that were in process when the error occurred will reprint
completely when the error is resolved. [SI-1]
User Action
1. Shut down the laser imager, then restart the imager.
2. If the error persists, call for service.
Show at DS
True
DICOM Printer
Status
FAILURE
Unique Symb. ID
ERR_PROCESSOR_RTD_OPEN
Logged for RMS: [TriggerValues=a, b][TriggerNotes=RTD zone number,
temperature in degrees C]
Logged for RMS
[TriggerValues=a, b][TriggerNotes=RTD zone number, temperature in
degrees C]