Adjustments and Replacements
9G9002 – 2011-08-15
2-29
Rollback Motor
Removal
[1]
Disconnect the rollback motor connector.
Important
An anti-rotational nut secures the lower screw that holds the motor. Press the nut against the back of the motor as
you untighten the lower screw to keep the nut from falling down when the motor is released.
[2]
Remove the:
(a)
2 screws
(b)
Motor
Installation
Reverse the steps in the removal procedure.
Prerequisites:
• Close the film cartridge. (Touch the Pause button on the display screen.)
• De-energize the laser imager.
• Disconnect the power cord.
•
Remove the right cover
Tools:
T20 Torx driver
H221_6006HC
Rollback
Motor