SERVICE MANUAL
4-8
9G9002 – 2011-08-15
Condition Codes
Subsystems in the laser imager
Note
Each subsystem has a range of error numbers. The numbers have two parts, a subsystem ID number
and an error ID number. This number system keeps the error ID numbers given to software
components in other laser imagers for similar errors. The subsystem ID and error number are only
separated by a hyphen on the Web Portal and not on the display screen or in the error log. For example,
the Web Portal displays “20-209” while the error log displays “20209.”
Subsystem IDs
Subsystem ID
Subsystem/Component Name
Error ID
MIM Core Components:
01
Configuration and Control Subsystem
01-xxx
04
DICOM Input Print Service SCP Subsystem
04-xxx
05
Study Reclamation Subsystem
05-xxx
06
Delivery Subsystem
06-xxx
DRE:
10
Delivery Subsystem (also contains certain system-wide
definitions)
10-xxx
13–14
Service tool or Web Portal User Interface
13-xxx,
14-xxx
MCS:
20
General MCS System
20-xxx
21
Film Supply Drawer
21-xxx
25
RF Tag
25-xxx
26
Film Transport (jams and film path controller errors)
26-xxx
27
Optics (beam power micro and FPGA)
27-xxx
28
Processor
28-xxx
29
Densitometer
29-xxx
36
Display Screen
36-xxx