7.3 REPLACEMENT PROCEDURES
Technical Service Manual
7 - 7
LifeCare PCA
7.3.3
FRONT
AND
REAR
ENCLOSURE
REPLACEMENT
The recommended tool for this procedure is a No. 2 Phillips screwdriver.
Replacement parts for this procedure are:
Enclosure, Front
Enclosure, Rear
Screw, 6-32 x 1/2, Pan Head, Phillips
Washer, Lock, #6
To replace the front and rear enclosures, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosures as described in
3. To replace the front enclosure, remove the specific components described
in
.
4. To replace the rear enclosure, remove the specific components described
in
.
5. Reassemble the replacement front or rear enclosure assembly (see
or
Note:
Verify all connectors are fully seated before the front and rear enclosures
are closed and fastened.
6. Join the front enclosure and rear enclosure in the exact reverse order of separation.
To verify successful replacement of the front and rear enclosures, perform the PVT
in