6.3 ALARM MESSAGES AND ERROR CODES
Technical Service Manual
6 - 9
LifeCare PCA
499
Alarm Data Error
Invalid alarm data
550
Power Down Error
Improper shutdown
551
Intertask Message Error
Illegal intertask message
552
Invalid Function Parameter
Invalid parameter data
553
Invalid Message
Invalid kernel message
554
Invalid Mode
Invalid user interface mode
555
Clock Change Error
Invalid date and time
during update
556
Invalid UI State Action
Invalid user interface
action code
557
Invalid Alarm
Illegal alarm message data
558
Mode Change Error
Illegal mode change
message
559
Invalid UI Stack
User interface stack
is corrupted or full
560
Drug Library Access Error
Invalid auto-programming
order or invalid CCA index
Check the drug library
against the order
588
Drug Table Data Error
Infuser time-out occurred
during installation of the drug
library
Power off infuser
Disconnect infuser from AC
power and remove battery;
then restart infuser
Reload drug library
If problem persists,
620
RAM Address Error
Address lines are stuck high,
stuck low, or shorted together
621
Chip ID Error
Register does not contain
expected value
622
Clock RAM Error
Data or address
communication errors
623
External RAM Error
Failed read/write operations
on external RAM
624
LCD Backlight
Backlight current
out of range
Verify LCD assembly
connections to MCU PWA
Replace LCD assembly
If problem persists,
Table 6-2. Error Codes Requiring Technical Service
Alarm
Code
Malfunction Description
Possible Cause
Corrective Action
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...