SECTION 7 REPLACEABLE PARTS AND REPAIRS
LifeCare PCA
7 - 20
Technical Service Manual
7.3.8.9
REMOVING
THE
CE
PWA
Note:
This procedure is intended to facilitate other repairs only.
The recommended tool for this procedure is a 3/16 nutdriver.
The replacement part for this procedure is:
Screw, 4/40 x 1/4, Hex Head, Slotted, with Washer
To remove the CE PWA, see
, and proceed as follows:
1. Disconnect the infuser from AC power.
2. Separate the front and rear enclosures as described in
3. Disconnect the antenna cable from the CE PWA.
4. Using the nutdriver, remove the four hex head screws that secure the retaining
bracket to the front enclosure. Remove the screws in the sequence shown
in
.
5. Grasp the PWA, moving it slightly away from the mechanism. Lift the PWA
with a rolling motion and remove it from the infuser.
6. Disconnect the ribbon cables from P4 and P19.
7. Reinstall the CE PWA and secure the retaining bracket to the front enclosure using
the screws that were removed in Step 4. Install the four hex head screws in the
sequence shown in
.
8. Reassemble the infuser in the exact reverse order of disassembly.
Note:
Verify all connectors are fully seated before the front and rear enclosures
are closed and fastened.
To verify successful installation of the CE PWA, perform the PVT in
and the network connectivity checks in
.
CAUTION:
The CE PWA is device specific and cannot be replaced with another
CE PWA. Infusion system communications will not function properly with a CE
PWA from another infuser. If the CE PWA is damaged, contact ICU Medical.
Содержание LifeCare PCA
Страница 2: ...LifeCare PCA Technical Service Manual This page intentionally left blank...
Страница 4: ...CHANGE HISTORY Technical Service Manual ii LifeCare PCA This page intentionally left blank...
Страница 8: ...CONTENTS LifeCare PCA vi Technical Service Manual Section 9 DRAWINGS 9 1 Appendix A 1...
Страница 22: ...SECTION 2 WARRANTY LifeCare PCA 2 2 Technical Service Manual This page intentionally left blank...
Страница 24: ...SECTION 3 SYSTEM OPERATING MANUAL LifeCare PCA 3 2 Technical Service Manual This page intentionally left blank...